Correction to: Variability in lutetium-177 SPECT quantification between different state-of-the-art SPECT/CT systems

Peters, Steffie M B, Meyer Viol, Sebastiaan L, van der Werf, Niels R, de Jong, Nick, van Velden, Floris H P, Meeuwis, Antoi, Konijnenberg, Mark W, Gotthardt, Martin, de Jong, Hugo W A M, Segbers, Marcel

DOI: https://doi.org/10.1186/s40658-021-00399-y

EJNMMI Physics 8 (1),

Abstract

Following publication of the original article [1], it was reported that the sphere volumes defined in the original article should be adjusted. The correct inner diameters (and volumes) of the spherical inserts were: 9.9mm (0.5 ml), 15.4mm (2.0 ml), 19.8 mm (4.0 ml), 24.8mm (8.0 ml), 31.3mm (16.0 ml) and 60mm (113 ml). Figures 3, 5 and 6 have been adjusted accordingly. The original article has been updated.